X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Evaluation of effective dose during abdominal three-dimensional imaging for three flat-panel-detector angiography
Shigeru Suzuki1, Ichiro Yamaguchi, Takashi Kidouchi
1Department of Radiology, Saitama Red Cross Hospital,Chuo-ku, Saitama, 338-8553, Japan. shig.suz@gmail.com
Abstract:
The purpose of this study was to evaluate the effective dose during abdominal three-dimensional (3D) imaging on phantoms and estimate the dose-area product (DAP) for effective dose conversion factors for three types of angiographic units. Three-dimensional imaging was performed for three sizes (small, medium, large) of human-shaped phantoms using three types of angiographic units (Allura Xper FD20/10, INNOVA 4100, AXIOM Artis dTA). We calculated 25 organ doses and effective doses using Monte Carlo technique for the three phantoms with a program for a personal computer. As benchmark studies to back up the results by Monte Carlo technique, we measured the organ doses directly on the small phantom using radiophotoluminescent glass dosimeters. The DAP value increased as the phantom size increased. The organ doses and the effective doses during the 3D imaging increased as the phantom size increased. The effective doses for the small phantom by Monte Carlo technique were 1.9, 2.2, and 2.1 mSv for the Allura Xper FD20/10, INNOVA 4100, and AXIOM Artis dTA, respectively, while those by direct measurement were 1.6, 2.0, and 2.6 mSv. The effective doses to DAP ratios by Monte Carlo technique were 0.37-0.45, 0.26-0.32, and 0.13-0.15 (mSv Gy⁻¹ cm⁻²) for the Allura Xper FD20/10, INNOVA 4100, and AXIOM Artis dTA, respectively. In conclusion, the effective doses during 3D imaging and the dose-to-DAP ratios differ among angiographic units, and the effective dose can be estimated using a proper conversion factor for each angiographic unit.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies VII: Vascular Imaging