Dosimetry for Cell Irradiation using Orthovoltage (40-300 kV) X-Ray Facilities
Published on: February 20, 2021
Dosimetry measurements with an intra-operative x-ray device
1Radiotherapy Physics, Royal Free Hospital, London, NW3 2QG, UK. davideaton@nhs.net
Abstract:
The INTRABEAM 50 kV x-ray device can be used for intra-operative partial breast irradiation. Spherical applicators are added to the source probe to deliver a radially symmetric radiation dose. Dosimetric data for calculation of absorbed dose were measured for this unit and a superficial unit with a similar beam quality, as defined by half value layer (HVL). Chamber calibration factors, N(K), and chamber correction factors, k(ch), were determined based on HVL, according to the IPEMB code of practice and addendum. Depth doses were also measured using an ionization chamber and GafChromic EBT film. HVL was measured as 0.85-1.30 mm Al across the range of applicator sizes. Values for N(K) and k(ch) were found to be similar for the two units and all INTRABEAM applicator sizes. Therefore, calibration of ionization chambers, radiochromic film and other relative dosimeters could be performed on the superficial unit. This has the advantage of higher dose rates and lower dependence on small variations in detector positioning. Depth dose measurements performed using film also agreed with chamber values, published and manufacturer data, giving a simple and robust method for commissioning and regular quality assurance.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...