Validation of the flexible electrogoniometer for measuring thoracic kyphosis
Diana M Perriman1, Jennifer M Scarvell, Andrew R Hughes
1Trauma and Orthopaedic Research Unit, Canberra Hospital and the Australian National University, Canberra, Australia. diana.perriman@anu.edu.au
Spine
|May 28, 2010
Summary
The flexible electrogoniometer (FEG) accurately measures thoracic kyphosis with excellent reliability. This tool shows strong correlation with the Cobb angle, offering a valid method for assessing spinal posture outside the lab.
Area of Science:
- Biomechanics
- Spinal health
- Medical device validation
Background:
- Thoracic kyphosis assessment traditionally relies on static Cobb angle measurements from radiographs.
- Existing non-invasive tools often provide only static measurements or are limited to laboratory settings.
- The flexible electrogoniometer (FEG) offers dynamic, out-of-laboratory joint angle measurement but requires validation for thoracic spine application.
Purpose of the Study:
- To evaluate the accuracy of the flexible electrogoniometer (FEG) for measuring thoracic kyphosis.
- To determine the test-retest reliability of the FEG in assessing thoracic spine angles.
- To establish the concurrent validity of the FEG by comparing its measurements to radiographic Cobb angles.
Main Methods:
- Bench testing of the FEG against a plurimeter to assess accuracy.
- Test-retest reliability assessment involving 12 subjects performing functional activities over two separate weeks.
- Concurrent validity evaluation by comparing FEG thoracic spine measurements with radiographic Cobb angles in upright and slumped postures.
Main Results:
- Excellent accuracy was observed between the FEG and plurimeter (r > 0.99), with minor loss at extreme ranges.
- Very strong test-retest reliability was demonstrated (intraclass correlation coefficient = 0.92).
- The FEG angle showed a high correlation with the mid-Cobb angle (r = 0.81), indicating good concurrent validity.
Conclusions:
- The flexible electrogoniometer (FEG) is an accurate and reliable tool for measuring thoracic kyphosis.
- FEG measurements demonstrate strong correlation with the gold-standard Cobb angle, particularly when aligned with the mid-Cobb angle.
- The FEG presents a promising, valid method for dynamic thoracic spine assessment in various settings.