The angular dependence of a 2-dimensional diode array and the feasibility of its application in verifying the
Qi-Lin Li1, Xiao-Wu Deng, Li-Xin Chen
1State Key Laboratory of Oncology in South China, Guangzhou, Guangdong 510060, PR China.
Background And Objective:
The planning dose distribution of intensity-modulated radiation therapy (IMRT) has to be verified before clinical implementation. The commonly used verification method is to measure the beam fluency at 0 degree gantry angle with a 2-dimensional (2D) detector array, but not the composite dose distribution of the real delivery in the planned gantry angles. This study was to investigate the angular dependence of a 2D diode array (2D array) and the feasibility of using it to verify the composite dose distribution of IMRT.
Methods:
Angular response of the central detector in the 2D array was measured for 6 MV X-ray, 10 cmx10 cm field and 100 cm source axis distance (SAD) in different depths. With the beam incidence angle of 0-60 degrees, at intervals of 10 degrees, and inherent buildup of the 2D array (2 g/cm2), the array was irradiated and the readings of the central diode were compared with the measurement of thimble ionization chamber. Using a combined 30 cmx30 cmx30 cm phantom which consisted of solid water slabs on top and underlying the 2D array, with the diode detectors placed at 8 g/cm2 depth, measurements were taken for beam angles of 0 degrees-180 degrees at intervals of 10 degrees and compared with the calculation of treatment planning system (TPS) that pre-verified with ion chamber measuring.
Results:
Differences between the array detector and thimble chamber measurements were greater than 1% and 3.5% when the beam angle was larger than 30 degrees and 60 degrees, respectively. The measurements in the combined phantom were different from the calculation as high as 20% for 90 degrees beam angle, 2% at 90 degrees+/-5 degrees and less than 1% for all the other beam angles.
Conclusions:
The 2D diode array is capable of being used in composite dose verification of IMRT when the beam angles of 90 degrees+/-5 degrees and 270 degrees+/-5 degrees are avoided.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Radiation: Applications
The average...
Modeling of Diode Forward Characteristics
Polar Coordinates: Problem Solving
Modeling of Diode Reverse Characteristics
When a reverse voltage applied to a Zener diode exceeds its breakdown voltage, the diode enters the breakdown region. At this point, the...