Design of a CMOS Potentiostat Circuit for Electrochemical Detector Arrays
Sunitha Ayers1, Kevin D Gillis, Manfred Lindau
1School of Electrical and Computer Engineering, Cornell University, Ithaca, NY. sb288@cornell.edu .
Abstract:
High-throughput electrode arrays are required for advancing devices for testing the effect of drugs on cellular function. In this paper, we present design criteria for a potentiostat circuit that is capable of measuring transient amperometric oxidation currents at the surface of an electrode with submillisecond time resolution and picoampere current resolution. The potentiostat is a regulated cascode stage in which a high-gain amplifier maintains the electrode voltage through a negative feedback loop. The potentiostat uses a new shared amplifier structure in which all of the amplifiers in a given row of detectors share a common half circuit permitting us to use fewer transistors per detector. We also present measurements from a test chip that was fabricated in a 0.5-microm, 5-V CMOS process through MOSIS. Each detector occupied a layout area of 35microm × 15microm and contained eight transistors and a 50-fF integrating capacitor. The rms current noise at 2kHz bandwidth is ≈ 110fA. The maximum charge storage capacity at 2kHz is 1.26 × 10(6) electrons.
Related Concept Videos
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential ensures...
Potentiometry: Overview
Potentiometry: Membrane Electrodes
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in the...
Controlled-Current Coulometry: Overview