Related Experiment Video
Updated: Jun 12, 2026

Optimized Griess Reaction for UV-Vis and Naked-eye Determination of Anti-malarial Primaquine
Published on: October 11, 2019
A new pre-column derivatization method for determination of gabapentin in human serum by HPLC using UV detection
Mohsen Amini1, Mohammad-Reza Rouini, Alireza Asad-Paskeh
1Department of Medicinal Chemistry and Pharmaceutical Sciences Research Center, Faculty of Pharmacy, Tehran University of Medical Sciences.
Abstract:
A high-performance liquid chromatography method using UV detection for determination of gabapentin in human serum has been developed. Serum was deproteinized with acetonitrile, and supernatant was evaporated. Derivatization was performed using 4-nitrophenylisothiocyanate, and separation was achieved on a C(18) column with detection at 335 nm. The method was linear in the 0.12-6.0 microg/mL range. No interference was found from endogenous amino acids. The method is suitable for routine therapeutic drug monitoring and for pharmacokinetic studies.
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
UV–Vis Spectrometers