Real-time stroke volume measurements for the optimization of cardiac resynchronization therapy parameters
José M Dizon1, T Alexander Quinn, Santos E Cabreriza
1Division of Cardiology, Department of Medicine, Columbia University Medical Center, New York, NY 10032, USA. jmd11@columbia.edu
Aims:
We investigated the utility of real-time stroke volume (SV) monitoring via the arterial pulse power technique to optimize cardiac resynchronization therapy (CRT) parameters at implant and prospectively evaluated the clinical and echocardiographic results.
Methods And Results:
Fifteen patients with ischaemic or non-ischaemic dilated cardiomyopathy, sinus rhythm, Class III congestive heart failure, and QRS >150 ms underwent baseline 2D echocardiogram (echo), 6 min walk distance, and quality of life (QOL) questionnaire within 1 week of implant. Following implant, 0.3 mmol lithium chloride was injected to calibrate SV via dilution curve. Atrioventricular (AV) delay (90, 120, 200 ms, baseline: atrial pacing only) and V-V delay (-80 to 80 ms in 20 ms increments) were varied every 60 s. The radial artery pulse power autocorrelation method (PulseCO algorithm, LiDCO, Ltd.) was used to monitor SV on a beat-to-beat basis (LiDCO, Ltd.). Optimal parameters were programmed and echo, 6 min walk, and QOL were repeated at 6-8 weeks post-implant. Nine patients had >5% increase in SV after optimization (Group A). Six patients had <5% improvement in SV (Group B). Compared with Group B, Group A had significant improvements in left ventricular ejection fraction (LVEF) (11.0 +/- 8.5 vs. 0.8 +/- 2.0%) and decrease in left ventricular end-diastolic dimension (LVEDD) (-0.6 +/- 0.4 vs. -0.2 +/- 0.2 cm) and 6 min walk (346 +/- 226 vs. 32 +/-271 ft, P < or = 0.05). Group A patients also tended to have greater improvement in the septal-to-posterior wall motion delay on M-mode echo (P = 0.07).
Conclusion:
Real-time SV measurements can be used to optimize CRT at the time of implant. Improvement in SV correlates with improvement in LVEF, LVEDD, and 6 min walk, and improvement in echocardiographic dyssynchrony.
Related Concept Videos
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Regulation of Stroke Volume
Preload refers to the degree of stretch on the heart before it contracts. It's analogous to the stretching of a rubber band; the more it's stretched, the more forcefully it snaps back. This concept is encapsulated in the Frank-Starling law of the...