X-ray and optical profiler analysis of electroformed x-ray optics
Abstract:
Electroformed grazing incidence mirrors were tested with x rays and were measured with an optical profiler. Their total effective areas are ?70% of the theoretical values both for 1.5- and 6.4-keV x rays. The half-power diameters (HPDs) are ?2 min of arc at 1.5 keV. Comparing the x-ray test result with the optical measurement, we set requirements of high-frequency roughness <14 A and a midfrequency roughness <70 A for a HPD <1.5 min of arc at 7 keV. We also present a design of the mirror array telescope for high energies (MARTHE) which has a total effective area of ~1000 cm(2) for 7- keV x rays.
More Related Videos
Related Concept Videos
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...