Related Experiment Video
Updated: Jun 12, 2026

Spatio-Temporal In Vivo Imaging of Ocular Drug Delivery Systems using Fiberoptic Confocal Laser Microendoscopy
Published on: September 27, 2021
Influence of container structures and content solutions on dispensing time of ophthalmic solutions
Keiji Yoshikawa1, Hiroshi Yamada
1Yoshikawa Eye Clinic, Tokyo, Japan. keiyomd@aol.com
Purpose:
To investigate the influence of container structures and content solutions on the time of dispensing from eye dropper bottles.
Methods:
Eye dropper bottle models, solution models (filtrate water/surfactant solution) and a dispensing time measuring apparatus were prepared to measure the dispensing time.
Results:
With filtrate water and pressure thrust load of 0.3 MPa, the dispensing time significantly increased from 1.1 +/- 0.5 seconds to 4.6 +/- 1.1 seconds depending on the decrease of inner aperture diameters from 0.4 mm to 0.2 mm (P < 0.0001). When using the bottle models with inner aperture diameters of 0.4 mm or larger, the dispensing time became constant. The dispensing time using surfactant solution showed the same tendency as above. When pressure thrust load was large (0.07 MPa), the solution flew out continuously with inner aperture diameters of 0.4 mm or larger and the dispensing time could not be measured. The inner aperture diameter most strongly explained the variation of the dispensing time in both the content solutions in the multiple linear regression analysis (filtrate water: 46%, R(2) = 0.462, surfactant solution: 56%, R(2) = 0.563).
Conclusions:
Among content solutions and container structures, the dispensing time was mostly influenced by the diameter of the inner aperture of bottles.
Related Concept Videos
Ophthalmic Drug Delivery Systems
Factors Influencing Drug Absorption: Pharmaceutical Parameters
Factors Affecting Dissolution: Drug Permeability, Stability and Stereochemistry
Compartment Models: Single-Compartment Model
Modified-Release Drug Delivery Systems: Influencing Factors
Formulation and Manufacturing Process: Physical Attributes of Generic Tablets and Capsules