Employing the Forced Oscillation Technique for the Assessment of Respiratory Mechanics in Adults
Published on: February 9, 2022
The Finometer can function as a standalone instrument in blood pressure variability studies and does not require
Michael K McMullen1, Julie M Whitehouse, Gillian Shine
1University of Westminster, London, UK. research@micmcmullen.se
Objective:
The Finometer records the beat-to-beat finger pulse contour and has been recommended for research studies assessing short-term changes of blood pressure and its variability. Variability measured in the frequency domain using spectral analysis requires the impact of breathing be restricted to high frequency spectra (>0.15 Hz) so that the data from participants need to be excluded when the breathing impact occurs in the low frequency spectra (0.04-0.15 Hz). This study tested whether breathing frequency can be estimated from standard Finometer recordings using either stroke volume oscillation frequency or spectral stroke volume variability maximum scores.
Methods:
Twenty-two healthy volunteers were tested for 270 s in the supine and upright positions. Finometer recorded the finger pulse contour and a respiratory transducer recorded breathing. Stoke volume oscillation frequency was calculated manually whereas the stroke volume spectral maximums were obtained using the software Cardiovascular Parameter Analysis. These estimates were compared with the breathing frequency using the Bland-Altman procedures.
Results:
Stroke volume oscillation frequency estimated breathing frequency to less than +/-10% and 95% levels of agreement in both supine (-7.7 to 7.0%) and upright (-6.7 to 5.4%) postures. Stroke volume variability maximum scores did not accurately estimate breathing frequency.
Conclusion:
Breathing frequency can be accurately derived from standard Finometer recordings using stroke volume oscillations for healthy individuals in both supine and upright postures. The Finometer can function as a standalone instrument in blood pressure variability studies and does not require support equipment to determine the breathing frequency.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: