Use of automated blood pressure measurements in clinical trials and registration studies: data from the VALTOP Study
Related Experiment Video
Updated: Jun 12, 2026

Software for Analysis of Heart Rate and Blood Pressure Time-series Data from the Valsalva Maneuver
Published on: June 27, 2025
Thomas Mengden1, Roland Asmar, Albert Kandra
1Centre of Vascular Medicine, Herz- und Gefäss-Campus Bad Nauheim, Germany. thomas.mengden@pitzer-kliniken.de
Background:
Auscultatory measurement of office blood pressure (BP) by mercury sphygmomanometers (AuscBPM) is still the gold standard in clinical trials and registration studies for antihypertensive drugs. The increasing availability of accurate automated oscillometric BP measuring devices has offered new perspectives in this field, although their usefulness in drug studies has not been systematically tested yet.
Methods:
During the course of Valsartan 320 mg EU Registration Study we used an electronic automated oscillometric BP measuring device (eBPM) as an alternative to conventional AuscBPM. Altogether 3776 patients were randomized into a double-blinded actively controlled parallel group study in 303 centers, and 54 422 BP readings were recorded by the validated, electronic, automated oscillometric device OMRON 705IT with digital printouts. Terminal digit preference and preference at therapeutic cutoff points were evaluated. The data were compared with the results of an earlier valsartan study similar in design but based on conventional AuscBPM. Furthermore, based on a simulation, four strategies for automated BP measurement with varying number of office readings (3-5) were analyzed to define an optimal method to collect BP at office visits.
Results:
eBPM eliminated terminal digit preference and dramatically reduced preferences for therapeutic cutoff points as compared with earlier valsartan trials with conventional AuscBPM. However, even with eBPM a minor bias with the therapeutic cutoff value was observed probably because of an observer bias during data documentation. The within-patient variability of three measurements sequentially taken at each visit was similar to other strategies including more measurements.
Conclusion:
On the basis of our data, we suggest that eBPM is a suitable alternative to AuscBPM in clinical trials and registration studies, and may carry specific advantages. Automatic data transfer of recorded values to electronic patient files may further minimize observer bias. Manufacturers should consider such findings for the development of professional devices.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.