Related Experiment Video
Updated: Jun 12, 2026

Quantifying Inferior Vena Cava Compliance and Distensibility in an In Vivo Ovine Model Using 3D Angiography
Published on: April 26, 2024
Difficult removal of retrievable IVC filters: a description of the "double-wire restraining" technique
Charles A Owens1, James T Bui, M- Grace Knuttinen
1Division of Interventional Radiology (M/C 931), Department of Radiology, University of Illinois Medical Center, 1740 West Taylor Street, Chicago, IL 60612, USA. cowens@uic.edu
Abstract:
We describe our experience with the use of the "double-wire restraining" technique to assist in the removal of two retrievable inferior vena cava filters: one had been misplaced in the right brachiocephalic vein with apex perforation of the vessel wall, and the second filter had migrated cephalad to straddle across both renal veins. The "double-wire restraining" technique consists of two stiff-shaft Glidewires (Terumo, Somerset, NJ) placed through the same introducer sheath and positioned on opposite sides of the filter. Both wires restrain the filter at the tip of the sheath as the sheath is advanced, thus allowing the operator to reposition the filter. This report details how this technique was used to realign two malpositioned filters and reposition the filter apices from their extravascular location, thus exposing them for ensnarement.