Related Experiment Video
Updated: Jun 12, 2026

Diffuse Reflectance Spectroscopy: Getting the Capillary Refill Test Under One's Thumb
Published on: December 2, 2017
Simultaneous measurement of X-ray specular reflection and off-specular diffuse scattering from liquid surfaces using
Yohko F Yano1, Tomoya Uruga, Hajime Tanida
1Research Organization of Science and Engineering, Ritsumeikan University, Shiga, Japan.
Abstract:
An X-ray reflectometer for simultaneous measurement of specular and off-specular reflection of liquid surfaces is described. The reflectometer, equipped with a two-dimensional single X-ray photon-counting pixel array detector obtained the full range of X-ray specular and off-specular reflections in an extremely short time (1 s). Both the specular and off-specular reflection of water exhibited good agreement with the predicted capillary-wave theory within the appropriate instrumental resolution. The approach is also demonstrated on an aqueous solution of 1-dodecyl-3-methylimidazolium chloride. The monolayer in which the dodecyl chain faces upwards and the Cl(-) anion locates next to the imidazolium ring formed on the water surface was found to be laterally homogeneous. The use of a pixel array detector will be particularly powerful for in situ measurements to investigate both out-of-plane and in-plane structures simultaneously, not only for liquid surfaces but also for other thin films.