Characterization and correction of beam-hardening artifacts during dynamic volume CT assessment of myocardial
Kakuya Kitagawa1, Richard T George, Armin Arbab-Zadeh
1Department of Medicine, Division of Cardiology, Johns Hopkins University School of Medicine, 1042 Ross Building, Baltimore, MD 21205, USA.
Purpose:
To fully characterize beam-hardening effects caused by iodinated contrast medium in the left ventricular (LV) cavity and aorta in the assessment of myocardial perfusion at computed tomography (CT) and to validate a beam-hardening artifact correction algorithm that considers fluid-filled vessels and chambers important sources of beam hardening.
Materials And Methods:
The Johns Hopkins University animal care and use committee approved all procedures. An anatomically correct LV and myocardial phantom to characterize beam-hardening artifacts was designed. Following validation in the phantom, the beam-hardening correction (BHC) algorithm was applied to 256-detector row dynamic volume CT images in a canine ischemia model (n = 5) during adenosine stress, and the effect of beam hardening was determined by comparing regional dynamic volume CT perfusion metrics (myocardial upslope normalized by maximum LV blood pool attenuation) with microsphere-derived myocardial blood flow (MBF). A paired Student t test was used to compare continuous variables from the same subject but under different conditions, while linear regression analysis was performed to estimate the slope and statistical significance of the relationship between CT-derived perfusion metrics and microsphere-derived MBF.
Results:
Beam-hardening artifacts were successfully reproduced in phantom studies and were eliminated with the BHC algorithm. The correlation coefficient of CT-derived perfusion metrics and microsphere-derived MBF improved from 0.60 to 0.74 (P > .05) following correction in the animal model.
Conclusion:
Beam-hardening artifacts confound dynamic volume CT assessment of myocardial perfusion. Application of the BHC algorithm is helpful for improving accuracy of myocardial perfusion at dynamic volume CT.
More Related Videos
12:15Tissue Preparation Techniques for Contrast-Enhanced Micro Computed Tomography Imaging of Large Mammalian Cardiac Models with Chronic Disease
Published on: February 8, 2022
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...