A comparative study of the effective radiation doses from cone beam computed tomography and plain radiography for
1Discipline of Oral and Maxillofacial Radiology, Faculty of Dentistry, University of Toronto, 124 Edward Street, Toronto, Ontario M5G 1G6, Canada.
Objectives:
As a first step in developing a protocol for multidimensional sialography using cone beam CT (CBCT), the objective of this study was to compare the effective radiation doses from sialography of the parotid and submandibular glands using plain radiography and CBCT.
Methods:
The effective doses were calculated from dose measurements made at 25 selected locations in the head and neck of a radiation analogue dosimeter (RANDO) phantom, using International Commission on Radiological Protection 2007 tissue weighting factors.
Results:
The effective dose (E) changed in relationship to changes in CBCT field of view (FOV), peak kilovoltage (kVp) and milliamperage (mA). Specifically, E decreased from a maximum of 932 microSv (30 cm FOV, 120 kVp, 15 mA) to 60 microSv (15 cm FOV, 80 kVp, 10 mA) for a parotid gland study and to 148 microSv (15 cm FOV, 80 kVp, 10 mA) for a submandibular study. The collective series of plain radiographs made during sialography of the parotid and submandibular glands yielded effective doses of 65 microSv and 156 microSv, respectively. The plain parotid gland series included one panoramic, two anterior-posterior skull and four lateral skull radiographs, whereas the submandibular gland series included one panoramic, one standard mandibular occlusal and four lateral skull radiographs.
Conclusion:
The effective doses from CBCT examinations centred on the parotid and submandibular glands were similar to those calculated for plain radiograph sialography when a 15 cm FOV was chosen in combination with exposure conditions of 80 kVp and 10 mA.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...