Contrast sensitivity comparison between AcrySof ReSTOR and Acri.LISA aspheric intraocular lenses
José F Alfonso1, Cristina Puchades, Luis Fernández-Vega
1Instituto Oftalmológico Fernández-Vega, Oviedo, Spain. j.alfonso@fernandez-vega.com
Journal of Refractive Surgery (Thorofare, N.J. : 1995)
|July 20, 2010
Summary
This study compared two intraocular lenses (IOLs): AcrySof ReSTOR and Acri.LISA. Both IOLs demonstrated normal contrast sensitivity in bright light but reduced sensitivity in dim light, with no significant differences between them.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Optics
Background:
- Intraocular lenses (IOLs) are crucial for vision restoration after cataract surgery.
- Assessing contrast sensitivity is vital for evaluating visual function under various lighting conditions.
- Aspheric IOL designs aim to improve visual quality, but their performance in mesopic conditions requires thorough evaluation.
Purpose of the Study:
- To compare the distance contrast sensitivity of AcrySof ReSTOR and Acri.LISA aspheric intraocular lenses (IOLs).
- To evaluate IOL performance under photopic, mesopic, and mesopic with glare conditions.
- To determine if significant differences exist between these two IOL models in contrast sensitivity post-implantation.
Main Methods:
- Contrast sensitivity was measured using the Optec 6500 FACT chart in patients implanted with either AcrySof ReSTOR SN6AD3 or Acri.LISA 366D IOLs.
- Testing was conducted under three lighting conditions: photopic (85 cd/m2), mesopic (3 cd/m2), and mesopic with glare.
- Binocular contrast sensitivity was assessed at 1 and 6 months post-surgery.
Main Results:
- Both AcrySof ReSTOR and Acri.LISA IOLs provided excellent distance and near visual acuity (approximately 20/20).
- Contrast sensitivity was within the normal range under photopic conditions for both IOLs.
- A reduction in contrast sensitivity, particularly at higher spatial frequencies, was observed under mesopic conditions for both IOLs, with no statistically significant differences between them.
Conclusions:
- The AcrySof ReSTOR SN6AD3 and Acri.LISA aspheric IOLs exhibit comparable performance in terms of contrast sensitivity.
- Both IOLs maintain normal contrast sensitivity in photopic conditions.
- A decrease in contrast sensitivity under mesopic conditions is noted for both IOLs, highlighting a common characteristic rather than a difference between the brands.
Related Concept Videos
Open Angle Glaucoma: Treatment
In open-angle glaucoma, the iridocorneal angle remains open, but the trabecular meshwork becomes stiff, slowing down the outflow of aqueous humor. This causes a buildup of aqueous humor in the anterior chamber, leading to a sudden increase in intraocular pressure. The treatment for open-angle glaucoma focuses on reducing the elevated intraocular pressure by either decreasing the secretion of aqueous humor or increasing its outflow.
Drugs such as carbonic anhydrase inhibitors, α2- and...
Drugs such as carbonic anhydrase inhibitors, α2- and...
Phase Contrast and Differential Interference Contrast Microscopy
Phase-Contrast Microscopes
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...