Comparing a traditional single optotype visual acuity test with a computer-based visual acuity test for childhood
Matthew B Schlenker1, Theodore J Christakis, Rosa M Braga-Mele
1Faculty of Medicine, University of Toronto, Toronto, Ont. matt.schlenker@gmail.com
Summary
A computer-based vision test (LazyeyeTest.org) showed higher sensitivity for detecting amblyopia in children than the traditional Sheridan Gardiner test (SGT). This free tool may improve screening efficiency and accuracy.
Area of Science:
- Ophthalmology
- Pediatric Vision Screening
- Diagnostic Technology Evaluation
Background:
- Traditional vision screening methods like the Sheridan Gardiner test (SGT) are widely used but may have limitations in detecting certain visual impairments.
- Computer-based tests (CBTs) offer potential advantages in standardization, engagement, and data collection for vision screening.
Purpose of the Study:
- To compare the effectiveness of a computer-based test (LazyeyeTest.org) against the Sheridan Gardiner test (SGT) for measuring visual acuity and detecting amblyopia in children.
- To evaluate the usability and efficiency of both vision screening methods.
Main Methods:
- A prospective, masked, cross-over study was conducted with kindergarten-aged children.
- Visual acuity was measured monocularly using both the SGT and LazyeyeTest.org.
- Referral rates, referral outcomes, and usability were compared between the two tests.
Main Results:
- The computer-based test (CBT) resulted in more low visual acuity scores, likely due to crowding bars on its optotypes.
- The CBT referred more children for further evaluation, with two cases of amblyopia identified among those referred.
- The CBT demonstrated higher sensitivity (88%) and specificity (92%) compared to the SGT, and was preferred by children and screeners for ease of use and engagement.
Conclusions:
- Replacing the SGT with LazyeyeTest.org could enhance the efficiency and sensitivity of amblyopia screening programs.
- Further research is recommended to validate these findings and assess the accuracy of lay screeners using the CBT.