Temperature-induced tissue susceptibility changes lead to significant temperature errors in PRFS-based MR thermometry

Sara M Sprinkhuizen1, Maurits K Konings, Martijn J van der Bom

  • 1Image Sciences Institute/Department of Radiology, University Medical Center Utrecht, Utrecht, The Netherlands. sara@isi.uu.nl

Summary

Temperature-dependent magnetic susceptibility in tissues causes significant errors in proton resonance frequency shift-based MR thermometry (MRT). This impacts the accuracy of temperature mapping, especially in fatty tissues like the breast.

Related Concept Videos

Temperature Measurement Sites01:14

Temperature Measurement Sites

A thermometer measures body temperature. The common sites for measuring body temperature are the oral cavity, axillary region, temporal artery, and skin surface, such as the forehead, abdomen, and axilla. True core body temperature is assessed in the rectum, tympanic membrane, pulmonary artery, esophagus, and urinary bladder.
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Factors Affecting Body Temperature01:28

Factors Affecting Body Temperature

As a nurse, it is vital to understand the factors affecting body temperature to monitor variations and effectively evaluate deviations from regular.
Factors may  include:
Assessing Body Temperature - Rectal01:27

Assessing Body Temperature - Rectal

Rectal temperature measurement is considered the most precise method for assessing core body temperature and typically registers higher than oral temperature. For adults, the rectal thermometer should be inserted 1 to 1.5 inches into the rectum to obtain the most accurate reading.
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Temporal Artery01:19

Assessing Body Temperature - Temporal Artery

Here is a stepwise guide to assessing the body temperature at the temporal artery using a temporal artery thermometer
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.  
Step 3: Assess the patient's forehead...
Equipments Used to Measure Body Temperature01:13

Equipments Used to Measure Body Temperature

Body temperature can be assessed using various devices and measured in Celsius or Fahrenheit.
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Decreased Body Temperature01:29

Decreased Body Temperature

A decreased body temperature can occur in patients with hypothermia and frostbite. Heat loss with extended cold exposure overpowers the body's ability to create heat, resulting in hypothermia. Core temperature readings help classify hypothermia. Mild hypothermia is temperatures between 32 °C (89.6 °F) and 35°C (95 °F) and is caused by impaired thermoregulation. Moderate hypothermia is temperatures between 28 C (82.4 °F) and 32 °C (89.6 °F) caused by sustained extreme cold exposure, and severe...