Contrast opacification using a reduced volume of iodinated contrast material and low peak kilovoltage in pulmonary CT
Andetta R Hunsaker1, Isabel B Oliva, Tianxi Cai
1Department of Radiology, Brigham and Women's Hospital and Harvard Medical School, Boston, MA 02115, USA. ahunsaker@partners.org
Objective:
The purpose of our study was to evaluate whether a reduced volume of iodinated contrast material for pulmonary CT angiography (CTA) using a low peak kilovoltage (kVp) technique yields equivalent opacification in all vessels.
Materials And Methods:
Four hundred fifty-two consecutive pulmonary CTA patients (265 women and 187 men; age range, 18-91 years; mean age, 56.2 years) were retrospectively evaluated. Patients were grouped into those receiving 125 mL (n = 229) and 75 mL (n = 223) of 370 mg I/mL iodinated contrast material. Low kVp was used in all patients. Hounsfield units were measured at lobar, posterobasal segment, posterobasal ramus, and medial basal subsegmental ramus in the left lower lobe. Three thoracic radiologists blinded to contrast dose independently and randomly evaluated the quality of enhancement using a 3-point scale at the same levels. The two-sample Student's t test was used to compare contrast opacification between groups; Spearman's correlation and the C-statistic were used to assess objective and subjective measurements. Interreader agreement was measured using Kendall's coefficient.
Results:
Mean contrast opacification differences between 125 mL versus 75 mL were not statistically significant (p > 0.21) at the lobar, segmental, and posterior basal rami, although the mean trended (p = 0.07) toward higher opacification of the medial basal ramus among 75 mL patients. Across all four pulmonary artery segments, there was good concordance between subjective and objective measurements, significantly higher than the null value of 0.50 (p > 0.05). For subsegmental arteries, concordance between objective and subjective measures was greater for the 75 mL group (p < 0.05). There was good interreader concordance, with a concordance coefficient of 0.70 (95% CI, 0.66-0.74).
Conclusion:
Both objective and subjective measures of contrast opacification support a reduction from 125 to 75 mL of contrast medium required for pulmonary CTA.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...