Related Experiment Video
Updated: Jun 10, 2026

Mouse Body Temperature Measurement Using Infrared Thermometer During Passive Systemic Anaphylaxis and Food Allergy Evaluation
Published on: September 14, 2018
A comparison of technologies used for estimation of body temperature
Jasdip Mangat1, Thomas Standley, Andrew Prevost
1Department of Medical Physics and Clinical Engineering, Cambridge University Hospitals NHS Foundation Trust, Cambridge, UK. jasdip.mangat@addenbrookes.nhs.uk
Abstract:
Body temperature measurement is an important clinical parameter. The performance of a number of non-invasive thermometers was measured by comparing intra- and inter-operator variability (n = 100) and clinical accuracy (n = 61). Variability was elevated in febrile compared to normothermic subjects for axillary and oral electronic contact thermometer measures and a temporal artery thermometer (p < 0.001 for both). Temporal artery thermometry and one mode of an infrared tympanic thermometer demonstrated significant clinical inaccuracy (p < 0.001 for both). Electronic contact thermometer repeatability and reproducibility are highly variable in febrile adults both in the axilla and oral cavity. Infrared thermometry of the skin over the superficial temporal artery is unreliable for measuring core body temperature, particularly in febrile subjects and patients in theatre. The infrared tympanic thermometers tested are acceptable for clinical practice; however, care should be exercised with the different modes of operation offered.
More Related Videos
04:54Using a Combination of Indirect Calorimetry, Infrared Thermography, and Blood Glucose Levels to Measure Brown Adipose Tissue Thermogenesis in Humans
Published on: June 2, 2023
08:22Using an Ingestible Telemetric Temperature Pill to Assess Gastrointestinal Temperature During Exercise
Published on: October 7, 2015
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...