Related Experiment Video
Updated: Jun 10, 2026

Ferromagnetic Bare Metal Stent for Endothelial Cell Capture and Retention
Published on: September 18, 2015
Fatigue life analysis and experimental verification of coronary stent
Jianjun Li1, Qiyi Luo, Zhiyong Xie
1MicroPort Medical (Shanghai) Co. Ltd, Shanghai 201203, PR China.
Abstract:
A computational and experimental method on biomechanics of stent is presented to analyze the stress distribution of different phases and evaluate the fatigue life according to Goodman criteria. As a result, the maximum stress and alternating stress were always located at the curvature area of rings, the fatigue bands in the experiment also verified the computation rationality. Matching between the numerical simulation and experimental results was satisfactory, which proved that the finite element analysis could provide theoretical evidence and help design and optimize the stent structure.
Related Concept Videos
Fatigue
Coronary Artery Disease V: Interprofessional Care
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...