Discrepancies in international normalized ratio results between instruments: a model to split the variation into
Una Ø Sølvik1, Per H Petersen, Grete Monsen
1Department of Public Health and Primary Health Care, Faculty of Medicine and Dentistry, University of Bergen, Norway. una.solvik@isf.uib.no
Differences in international normalized ratio (INR) measurements between devices are significant. Patient-specific factors contribute to variations, recommending consistent use of a single INR device per patient for reliable monitoring.
Area of Science:
- Clinical Chemistry
- Medical Devices
- Hematology
Background:
- Significant variations in international normalized ratio (INR) measurements between different instruments exceed expected analytical imprecision.
- Understanding the sources of these discrepancies is crucial for accurate patient management.
Purpose of the Study:
- To quantify the subcomponents of variation contributing to differences in INR measurements across multiple devices.
- To investigate the impact of patient-specific factors on INR measurement variability.
Main Methods:
- Compared INR measurements from one laboratory and three point-of-care instruments using blood samples from 36 warfarin-treated outpatients.
- Employed linear regression to correct for systematic deviations and ANOVA for nested design to partition variation into between-subject, within-subject, and analytical components.
Main Results:
- Mean differences between instruments ranged from 1.0% to 14.3%.
- Between-subject variation (CV) ranged from 3.3% to 7.4%, while within-subject variation was consistently around 5%.
- Analytical imprecision varied between 3.8% and 8.6%.
Conclusions:
- INR measurement differences are attributable to calibration, patient-specific variations (between- and within-subject), and analytical imprecision.
- Patient plasma effects likely drive between- and within-subject variations, causing INR differences to fluctuate over time for individual patients.
- Consistent use of a single INR instrument type per patient and site is recommended to minimize monitoring variability.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Variability: Analysis
The range is a simple measure of variability, indicating the difference between the highest and...
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
One-Compartment Open Model: Wagner-Nelson and Loo Riegelman Method for ka Estimation
On...
Uncertainty in Measurement: Reading Instruments