Background:
Ventilator frequency is one of the determinants of tidal volume delivery during high-frequency ventilation. Clinicians increasingly use data on ventilator displays to inform their decisions.
Aim:
To measure the frequencies delivered by the Dräger Babylog 8000plus ventilator when used in high-frequency mode.
Methods:
Ventilator waveforms using a test lung were recorded at the full range of settings 5-20 Hz using Spectra software at 1000 Hz. The changes in frequency produced by a 1-Hz change in set frequency were calculated. Actual and displayed frequencies were compared.
Results:
For settings up to 12 Hz, median (range) difference between set and delivered frequencies was 0 (-0.4 to +0.1) Hz. Above 12 Hz, delivered frequency varied by -0.3 (-1.9 to +0.3) Hz. For 1-Hz changes in frequency settings, in the range 5-12 Hz, 1-Hz changes produced a change in delivered frequency of 1.0 (0.6-1.4) Hz. Above 12 Hz, the corresponding changes were 0.7 (0-2.9) Hz. The ventilator displays the set frequency during operation rather than the delivered frequency.
Conclusion:
At 12 Hz and below, the differences between set and delivered frequencies were relatively small compared with those at 13 Hz and higher. Above 13 Hz, the difference between set and delivered frequencies was up to 2.9 Hz. Some frequency setting changes did not result in a change in delivered frequency.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...