Related Experiment Video
Updated: Jun 9, 2026

Tissue-simulating Phantoms for Assessing Potential Near-infrared Fluorescence Imaging Applications in Breast Cancer Surgery
Published on: September 19, 2014
Novel spherical phantoms for Q-ball imaging under in vivo conditions
Amir Moussavi-Biugui1, Bram Stieltjes, Klaus Fritzsche
1Department of Medical Physics in Radiology, German Cancer Research Center, Heidelberg, Germany.
Abstract:
For the validation of complex diffusion imaging techniques like q-ball imaging that aim to resolve multiple fiber directions, appropriate phantoms are highly desirable. However, previous q-ball imaging phantoms had diffusion anisotropies well below those of in vivo white matter. In this work, fiber phantoms of well-defined geometry are presented. The fibers are wound on a spherical spindle yielding high packing densities and consequently high diffusion anisotropies (fractional anisotropy 0.93 ± 0.02 at b = 500 s/mm(2)). Phantoms with 90° and 45° crossing angle were constructed both with two crossing types. In the "stacked" crossing, two fiber strings were wound consecutively to simulate two touching fibers, in the "interleaved" crossing, fibers were wound alternately. The stacked crossing allows the alteration of partial volumes, whereas the interleaved crossing provides constant partial volumes, allowing e.g. the easy alteration of the SNR by varying the slice thickness. Exemplary q-ball imaging validation measurements using different b-values and slice thicknesses are presented.