Related Experiment Video
Updated: Jun 9, 2026

Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
In vivo temperature measurement during transcatheter endomyocardial Nd-YAG laser irradiation in dogs
H P Weber1, A Heinze, G Hauptmann
1Laser and Electrophysiology Laboratory, Department of Medicine, Harlaching Hospital, Teaching Hospital of the University of Munich, Munich, Germany.
Abstract:
Intramural temperatures were monitored during catheter-directed endomyocardial non-contact laser irradiation by means of an array of thermocouples inserted epicardially. Via a novel electrode-laser catheter, a total of 30 Nd-YAG laser pulses, 1064 nm, 25 W/15 s, 20 W/30 s, 15 W/45 s,n = 10 each, were aimed at various endocardial sites in the ventricles of four dogs. Temperatures increased gradually after the onset of the laser pulses. Maximum values were measured in the central area of the irradiated spots: 100 +/- 15 double daggerC at 25 W, 78+/- 23 double daggerC at 20 W and 80 +/- 13 double daggerC at 15 W. Application of 300 J of laser energy induced significantly higher temperatures at 25 W (p< 0.02) than at 20 or 15 W (20vs 15 W: p> 0.05). Initial temperature rise and lesion formation was accompanied by dwindling of local electrical potentials. There was no crater formation. Transcatheter non-contact Nd-YAG laser coagulation of healthy ventricular myocardium can be performed in a safe and controllable manner at power settings up to 20 W and laser energies up to 675 J.
More Related Videos
06:45In Vitro and In Vivo Delivery of Magnetic Nanoparticle Hyperthermia Using a Custom-Built Delivery System
Published on: July 2, 2020
09:23Inducing Targeted Mild Hyperthermia in Murine Tumor Models through Photothermal Conversion of Near-infrared Light by Intratumoral Gold Nanorods
Published on: October 10, 2025