Sequential dual-energy subtraction technique with a dynamic flat-panel detector (FPD): primary study for image-guided
Rie Tanaka1, Shigeru Sanada, Takeshi Matsui
1Division of Health Sciences, Radiology, Graduate School of Medical Science, Kanazawa University, Kodatsuno, Kanazawa, Japan. rie44@mhs.mp.kanazawa-u.ac.jp
Abstract:
A sequential dual-energy subtraction technique for image-guided radiation therapy (IGRT) was developed. Here, we report on a computerized method for creating sequential soft-tissue images and the accuracy of tracking targets on the images obtained, in comparison to conventional fluoroscopic images. Two sets of sequential chest images during respiration of a normal subject were obtained with X-rays of different energy separately with a flat-panel detector (FPD). Sequential soft-tissue images were created from the two sets of sequential images consisting of real-time images and reference template images, respectively. The creation of sequential soft-tissue images consisted of three steps: one-to-one image correspondence of the two sequential images, image registration, and image subtraction in each frame. Motion tracking of lung vessels was then performed by the template-matching technique. For evaluation of the accuracy of motion tracking on the sequential soft-tissue images, the results were compared with those on the original sequential images. Sequential soft-tissue images provided more accurate tracking than the original images (P < 0.01). There was no significant error throughout all frames in the soft-tissue images, whereas the rib shadow introduced a tracking error in the original images. The maximum errors were 4.1 +/- 0.3 mm in the sequential soft-tissue images and 28.1 +/- 20.0 mm in the original images. In conclusion, sequential soft-tissue images were helpful for tracking of a target affected by respiratory motion. Dual-energy subtraction has the potential to improve the accuracy of IGRT without implanted markers.
More Related Videos
09:34Dynamic Imaging of Chimeric Antigen Receptor T Cells with [18F]Tetrafluoroborate Positron Emission Tomography/Computed Tomography
Published on: February 17, 2022
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...