Related Experiment Video
Updated: Jun 9, 2026

Rapid Formation and Testing of Self-expanding NiTi Frames with a Small Form Factor Suitable for Minimally Invasive Implants
Published on: March 7, 2025
Synthesis and characterisation of a new superelastic Ti-25Ta-25Nb biomedical alloy
E Bertrand1, T Gloriant, D M Gordin
1INSA Rennes, UMR CNRS 6226 SCR/Chimie-Métallurgie, 20 avenue des Buttes de Coësmes, 35043 Rennes Cedex, France.
Abstract:
In this study, a new Ti-25Ta-25Nb (mass%) beta alloy was synthesised by cold crucible semi-levitation melting. This technique made it possible to obtain homogeneous ingots although the elements used have very different melting points. After melting, a thermo-mechanical treatment was applied in order to obtain a perfectly recrystallised beta microstructure. For this alloy composition, the tensile tests showed a very low Young's modulus associated with an important super-elastic behaviour, which contributes to decrease the elastic modulus under stress and to increase the recoverable strain. On the other hand, the corrosion tests, which were carried out in a neutral Ringer solution, indicated a corrosion resistance higher than that of the commercially pure CP Ti alloy. These results show that this new alloy possesses all the characteristics necessary for its long-term use in medical implants.