Dosimetric characterization of the iBEAM evo carbon fiber couch for radiotherapy
David W Smith1, Damianos Christophides, Christopher Dean
1Department of Medical Physics and Clinical Engineering, St. James's Institute of Oncology, Bexley Wing, St. James's University Hospital, Leeds LS9 7TF, United Kingdom. David-W.Smith@leedsth.nhs.uk
Purpose:
This study characterizes the dosimetric properties of the iBEAM evo carbon fiber couch manufactured by Medical Intelligence and examines the accuracy of the CMS XiO and Nucletron Oncentra Masterplan (OMP) treatment planning systems for calculating beam attenuation due to the presence of the couch.
Methods:
To assess the homogeneity of the couch, it was CT scanned at isocentric height and a number of signal intensity profiles were generated and analyzed. To simplify experimental procedures, surface dose and central axis depth dose measurements were performed in a solid water slab phantom using Gafchromic film for 6 and 10 MV photon beams at gantry angles of 0 degree (normal incidence), 30 degrees, and 60 degrees with an inverted iBEAM couch placed on top of the phantom. Attenuation measurements were performed in a cylindrical solid water phantom with an ionization chamber positioned at the isocenter. Measurements were taken for gantry angles from 0 degree to 90 degrees in 10 degrees increments for both 6 and 10 MV photon beams. This setup was replicated in the XiO and OMP treatment planning systems. Dose was calculated using the pencil beam, collapsed cone, convolution, and superposition algorithms.
Results:
The CT scan of the couch showed that it was uniformly constructed. Surface dose increased by (510 +/-0)% for a 6 MV beam and (600 +/- 20)% for a 10 MV beam passing through the couch at normal incidence. Obliquely incident beams resulted in a higher surface dose compared to normally incident beams for both open fields and fields with the couch present. Depth dose curves showed that the presence of the couch resulted in an increase in dose in the build up region. For 6 and 10 MV beams incident at 60 degrees, nearly all skin sparing was lost. Attenuation measurements derived using the ionization chamber varied from 2.7% (0 degree) to a maximum of 4.6% (50 degrees) for a 6 MV beam and from 1.9% (0 degree) to a maximum of 4.0% (50 degrees) for a 10 MV beam. The pencil beam and convolution algorithms failed to accurately calculate couch attenuation. The collapsed cone and superposition algorithms calculated attenuation within an absolute error of +/- 1.2% for 6 MV and +/- 0.8% for 10 MV for gantry angles from 0 degree to 40 degrees. Some differences in attenuation were observed dependent on how the couch was contoured.
Conclusions:
These results demonstrate that the presence of the iBEAM evo carbon fiber couch increases the surface dose and dose in the build up region. The inclusion of the couch in the planning scan is limited by the field of view employed and the couch height at the time of CT scanning.
More Related Videos
Related Concept Videos
Absorption of Radiation
Biological Effects of Radiation
Radiation: Applications
The average...
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...