Erythrocyte Sedimentation Rate: A Physics-Driven Characterization in a Medical Context
Published on: March 24, 2023
Evaluation of the Ves-Matic Cube 200 erythrocyte sedimentation method: comparison with Westergren-based methods
Joyce Curvers1, Jurgen Kooren, Maartje Laan
1Clinical Laboratory, Catharina Hospital Eindhoven, PO Box 1350, 5602 ZA Eindhoven, the Netherlands.
Abstract:
The erythrocyte sedimentation rate (ESR) is still a widely used parameter for acute phase inflammation. Recently, new methods based on direct undiluted measurement of ESR in a standard EDTA tube have been developed. We evaluated the analytic performance of one of these new methods, the Ves-Matic Cube 200 (Diesse Diagnostica Senese, Siena, Italy), and compared it with several established Westergren-based diluted methods. The Ves-Matic Cube 200 showed a poor correlation (r = 0.83) with the International Council for Standardization in Haematology Westergren reference method, mainly caused by a considerable negative bias at low ESR levels. Moreover, a random bias was found at higher ESR levels that correlated with hematocrit levels, suggesting a differential influence of packed cell volume on the Ves-Matic Cube 200 results compared with Westergren results. We conclude that the Ves-Matic Cube 200 method is not interchangeable with Westergren-based diluted methods and generates ESR results that are too deviant to be clinically acceptable.