The dependency of air caloric stimulus effectiveness on delivery tip characteristics
1Department of Medical Physics & Clinical Engineering, Royal Liverpool University Hospital, Liverpool, UK. g.lightfoot@liverpool.ac.uk
Abstract:
This brief study investigates the extent and origin of the apparent dependency of air delivery speculum size on the effectiveness of air used as a stimulus in the bi-thermal caloric test, using water as a reference stimulus. Eleven normal volunteers served as subjects. Six caloric stimulus delivery conditions included air with two speculum sizes, speculum only and with a rubber tube extension, and water. The resulting nystagmus was used as an index of stimulus efficiency. The effectiveness of the air stimulus was found to be dependent on the diameter of the speculum used for air delivery. The narrower speculum generated nystagmus that was typically twice that generated by the wider speculum. Users of air caloric equipment whose design includes a speculum that influences the effectiveness of the stimulus should be aware of this dependency and ensure their clinical interpretation of results is made with reference to appropriate normative data.
Related Concept Videos
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Constant Pressure Calorimetry
Joule-Thomson Effect
This experiment forces high-pressure gas through a throttle valve or a porous plug to a lower-pressure region. The gas expands as it passes through to...
Bernoulli's Principle: Applications
Entrainment devices use a high fluid speed to create low pressures and, thus, entrain one fluid into another. Some examples of these devices are given below:
Other Factors Affecting Respiration Centers
However, the ability to hold one's breath voluntarily is not limitless. When the CO2 concentration in the blood reaches a critical level,...