Dynamic contrast-enhanced CT studies: balancing patient exposure and image noise
Gunnar Brix1, Ursula Lechel, Markus Petersheim
1Department of Medical and Occupational Radiation Protection, Federal Office for Radiation Protection, Oberschleissheim, Germany. gbrix@bfs.de
Objectives:
To establish the essential basis for balancing the dose versus noise trade-off in dynamic contrast-enhanced (DCE) CT by means of a phantom study.
Materials And Methods:
Measurements were performed at a 64-section dual-source system, using the default protocols for DCE imaging (40 scans) of the trunk (current-time product per scan, 100 mAs; voltage, 120 kVp; pixel size, 0.9 × 0.9 × 8 mm3; CTDIvol per examination, 264 mGy) and head (270 mAs, 80 kVp, 0.45 × 0.45 × 8 mm3, 429 mGy). For 3 representative sections of an anthropomorphic phantom (head, upper abdomen, pelvis) transaxial dose distributions were measured by thermoluminescent dosimeters. The image noise was determined for 5 values of the current-time product (but otherwise identical parameter settings) and 4 pixel resolutions at a water-filled trunk and head phantom.
Results:
Highest exposures occurred at the periphery of the trunk and head with maximum skin entrance doses of about 300 mGy. Effective doses related to the 3 exposure scenarios were between 4 and 20 mSv, but were not at all predictive of local exposure levels. The image noise was inversely proportional to the square root of the current-time product and, with restrictions, to the pixel size. Noise levels determined for the standard settings were 13.8 HU (trunk) and 4.4 HU (head) and thus comparable with the contrast enhancement typically detected in tumors and ischemic brain tissues, respectively.
Conclusions:
The opposing requirements of risk and noise limitation in DCE-CT cannot be balanced without substantially reducing the spatial resolution. But even so, local radiation exposures are rather high for a diagnostic procedure. Indications to perform a DCE examination should thus be strictly limited to patients who really benefit from it.
More Related Videos
08:02Novel Quantification Protocol for Cardiovascular Calcification Progression Using Longitudinal MicroPET/MicroCT Images
Published on: November 15, 2024
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...