Absolute efficiency and time-response measurement of single-photon detectors
Abstract:
Using correlated photons from spontaneous parametric downconversion, we have measured both the absolute quantum efficiencies and the time responses of four single-photon detectors. Efficiencies as high as (76.4 ± 2.3)% (at 702 nm) were seen, which to our knowledge are the highest reported single-photon detection efficiencies. An auxiliary retroreflection mirror was found to increase the net detection efficiency by as much as a factor of 1.19. The narrowest time profile for coincidences between two detectors displays a peak with 300 ps FWHM. We also investigated the presence of afterpulses and the effects of saturation and varying device parameters.