Deformations of overloaded bands under pH-stable conditions in reversed phase chromatography
Lena Edström1, Jörgen Samuelsson, Torgny Fornstedt
1Department of Physical and Analytical Chemistry, Uppsala University, BMC Box 599, SE-751 24 Uppsala, Sweden.
Abstract:
It has recently been demonstrated, using mathematical models, how peculiar overloaded band profiles of basic compounds are due to the local pH in the column when using low capacity buffers. In this study, overloaded peak shapes resulting after injection of carefully pH matched samples close to the pK(a) of the chosen solute are investigated primarily on two columns; one hybrid silica C18 column (Kromasil Eternity) and one purely polymeric column (PLRP-S), the latter lacking C18 ligands. It was found that distorted peaks of the basic test compound appear even though there is no difference in pH between the injected sample solution and the eluent; the previous explanation to why these effects occur is based on a pH mismatch. Thus, the unusual band shapes are not due to an initial pH difference. Furthermore, it was observed that the effect does not appear on polymeric columns without C18 ligands, but only on columns with C18 ligands, independently of the base matrix (silica, hybrid silica, polymeric).
More Related Videos
05:45Capillary Electrophoresis-based Hydrogen/Deuterium Exchange for Conformational Characterization of Proteins with Top-down Mass Spectrometry
Published on: June 8, 2021
05:52Histone Modification Screening using Liquid Chromatography, Trapped Ion Mobility Spectrometry, and Time-Of-Flight Mass Spectrometry
Published on: January 12, 2024
Related Concept Videos
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Chromatographic Methods: Terminology
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Chromatographic Resolution
The effectiveness of separation can be evaluated by determining the level of separation between two neighboring peaks in a chromatogram, which represents the individual components of a sample.
In chromatography,...