Sterility maintenance assessment of moist/wet material after steam sterilization and 30-day storage
Giovana Abrahão de Araújo Moriya1, Kazuko Uchikawa Graziano
1Hospital Alemão Oswaldo Cruz, Universidade de Sao Paulo, Brasil. gaamoriya@gmail.com
Abstract:
Moist/wet materials stored after autoclaving are considered contaminated and not recommended for use. This study evaluates the maintenance of sterility in moist/wet material after being submitted to steam sterilization and stored for a period of 30 days. Aiming to support decision-making in emergency situations, 40 surgical boxes packed in nonwoven cloth covering Spunbound, Metblouwn, Spunbound (SMS): half (the experimental group) were placed in an autoclave but the drying phase was interrupted, yielding moist/wet materials and the other half (the negative control group) underwent the complete cycle. The external parts of each surgical box were deliberately contaminated with Serratia marcescens and subsequently stored for 30 days. After this period, the boxes' contents were submitted to sterility tests and no growth was observed. The presence of moisture inside the surgical boxes did not interfere with maintaining their sterility.
Insights
Sterile processing of medical supplies can be challenging. This study found that moist materials, even when contaminated externally, remained sterile for 30 days after autoclaving, supporting their use in emergencies.
Area of Science:
- Microbiology
- Sterilization Science
- Healthcare Management
Background:
- Moist materials after autoclaving are typically deemed contaminated.
- This poses challenges for sterile supply management, especially in emergencies.
Purpose of the Study:
- To evaluate the sterility of moist/wet materials after steam sterilization and 30-day storage.
- To inform decision-making regarding the use of such materials in critical situations.
Main Methods:
- 40 surgical boxes in Spunbound, Metblouwn, Spunbound (SMS) nonwoven cloth were used.
- Experimental group: autoclave cycle interrupted before drying (moist).
- Control group: complete autoclave cycle (dry).
- External contamination with Serratia marcescens.
- 30-day storage followed by sterility testing.
Main Results:
- No microbial growth was observed in the contents of any surgical box.
- External contamination did not compromise internal sterility.
- Moisture presence did not affect the maintenance of sterility.
Conclusions:
- Moist/wet materials stored after autoclaving maintained sterility for 30 days.
- This finding challenges the conventional view of moist materials as contaminated.
- Supports the potential use of moist sterile packs in emergency scenarios.
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Key Techniques in Microbiology
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Ostomy Care
An ostomy is a surgical procedure that creates an artificial opening from the intestines to the outside of the body, allowing for the rerouting of effluent. This opening is known as a stoma. A stoma usually protrudes above the skin surface, appearing pink or red, moist, and round, and it lacks nerve sensations.
There are different types of ostomies, including colostomies, ileostomies, and urostomies:
Physical Methods for Controlling Microbial Growth: Temperature