Related Experiment Video
Updated: Jun 8, 2026

Extraction of Non-Protein Amino Acids from Cyanobacteria for Liquid Chromatography-Tandem Mass Spectrometry Analysis
Published on: December 9, 2022
Automated protein hydrolysis delivering sample to a solid acid catalyst for amino acid analysis
1Biomolecular Characterization Team, Riken Advanced Science Institute, and CREST, JST, 2-1 Hirosawa, Wako, Saitama 351-0198.
Abstract:
In this study, we developed an automatic protein hydrolysis system using strong cation-exchange resins as solid acid catalysts. Examining several kinds of inorganic solid acids and cation-exchange resins, we found that a few cation-exchange resins worked as acid catalysts for protein hydrolysis when heated in the presence of water. The most efficient resin yielded amounts of amino acids that were over 70% of those recovered after conventional hydrolysis with hydrochloric acid and resulted in amino acid compositions matching the theoretical values. The solid-acid hydrolysis was automated by packing the resin into columns, combining the columns with a high-performance liquid chromatography system, and heating them. The amino acids that constitute a protein can thereby be determined, minimizing contamination from the environment.