Near Infrared Optical Projection Tomography for Assessments of β-cell Mass Distribution in Diabetes Research
Published on: January 12, 2013
Contrast and resolution analysis of iterative angular domain optical projection tomography
Eldon Ng1, Fartash Vasefi, Bozena Kaminska
1Imaging Program, Lawson Health Research Institute, 268 Grosvenor St. London, Ontario, Canada.
Image contrast and resolution in Angular Domain Imaging (ADI) depend on location. This study characterized ADI performance, finding iterative reconstruction superior to back projection for artifact reduction.
Area of Science:
- Medical imaging
- Image reconstruction algorithms
- Angular Domain Imaging (ADI)
Background:
- Image quality in Angular Domain Imaging (ADI) is known to be position-dependent.
- Understanding these variations is crucial for accurate image interpretation and system optimization.
Purpose of the Study:
- To quantitatively characterize the contrast and resolution of an ADI system across multiple locations in the imaging plane.
- To compare the performance of filtered back projection (FBP) and iterative reconstruction (IR) algorithms in reconstructing various targets within the ADI system.
Main Methods:
- Systematic evaluation of contrast and resolution at diverse positions within the ADI imaging plane.
- Comparative analysis of image reconstructions generated by FBP and IR algorithms.
- Assessment of artifact robustness for both reconstruction methods.
Main Results:
- Contrast exhibited significant variation with depth but minimal change with lateral position.
- Resolution demonstrated significant dependence on lateral position with minimal variation concerning depth.
- The iterative reconstruction algorithm proved robust against common artifacts like ring and streak artifacts.
- Filtered back projection reconstructions were susceptible to artifacts stemming from incomplete projection data.
Conclusions:
- Positional variations in contrast and resolution are inherent characteristics of ADI systems.
- Iterative reconstruction offers superior artifact management compared to filtered back projection for ADI.
- The choice of reconstruction algorithm significantly impacts image quality and artifact levels in ADI.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...