Design of an open path near-infrared diode laser sensor: application to oxygen, water, and carbon dioxide vapor
Abstract:
An open path diode laser sensor was constructed with near-infrared diode lasers and two-tone frequency-modulation spectroscopy. The sensor incorporates several novel features (such as digital signal-processing algorithms, a computerized line-locking routine, and discontinuous wavelength scanning) that are important in a field instrument. The sensor was used to monitor oxygen, water, and carbon dioxide in the near-infrared spectral range. For oxygen, an absorbance detection sensitivity of 2 × 10(-6) in a 10-Hz bandwidth was demonstrated with a GaALAs laser at 760.56 nm. The stability of the sensor was 0.1% over a period of 10 h when an absorbance of 6 × 10(-3) was monitored.
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...