Related Experiment Video
Updated: Jun 8, 2026

Reduced-gravity Environment Hardware Demonstrations of a Prototype Miniaturized Flow Cytometer and Companion Microfluidic Mixing Technology
Published on: November 13, 2014
Miniaturization of a magnetically levitated axial flow blood pump
Shanbao Cheng1, Mark W Olles, Don B Olsen
1Department of Mechanical Engineering, Rochester Institute of Technology, Mechanical Engineering, Rochester, NY 14623, USA.
Abstract:
This article introduces a unique miniaturization process of a magnetically levitated axial flow blood pump from a functional prototype to a pump suitable for animal trials. Through COMSOL three-dimensional finite element analysis and experimental verification, the hybrid magnetic bearings of the pump have been miniaturized, the axial spacing between magnetic components has been reduced, and excess material in mechanical components of the pump was reduced. Experimental results show that the pump performance was virtually unchanged and the smaller size resulted in the successful acute pump implantation in calves.