Use of independent component analysis for reducing CPR artefacts in human emergency ECGs
Marcus Granegger1, Tobias Werther, Hermann Gilly
1Department of Anesthesia, Intensive Care Medicine and Pain Therapy, Medical University of Vienna, Vienna, Austria. marcus.granegger@meduniwien.ac.at
Aim:
We applied independent component analysis (ICA) to cardiopulmonary resuscitation (CPR)-corrupted human multichannel emergency ECGs with the aim of reconstructing the original ECGs.
Materials:
Two ICA algorithms (EFICA and JADE) were selected. Data for ICA were acquired by simultaneously recording eight ECG channels during CPR on a porcine model. The algorithms' reconstruction performance was assessed by the Spearman correlation coefficient (SCC) and the shock advice algorithm of an AED. We then compared the performance of EFICA with the established second-channel adaptive matching pursuit method (AF).
Results:
ICA was applied to 918 corrupted ECG multichannel signals. The sensitivity of the AED's shock/no-shock decision increased from 93.5% (corrupted signal) to 99.5/99.8% (JADE/EFICA) in the selected independent component; specificity increased from 50.5% to 78.9/83.2% (JADE/EFICA). The SCCs comparing the reconstructed with the original signal (JADE: 0.75±0.15; EFICA: 0.76±0.15, n=918) were significantly higher than for the corrupted signal vs. the original (0.52±0.22). The SCC is significantly higher (p<0.01) using EFICA than AF (EFICA: 0.75±0.16; AF: 0.72±0.19, n=718). For all signals at all SNR levels, specificity did not differ significantly between EFICA (83.6%) and AF (80.2%). EFICA proved to be superior especially at low corruption levels (SNR<-5dB). Sensitivity was above 99.5% for both algorithms.
Conclusion:
We have demonstrated that CPR artefacts in the emergency ECG can be reduced using ICA. EFICA and JADE are at least as successful in this regard as are other published algorithms. In particular, non-shockable signals with low SNRs (<-5dB) are reconstructed significantly better (p=0.01) with EFICA than with AF.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Cardiopulmonary Resuscitation III: AED Use
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...