Automated cerebrospinal fluid cell counts using the Sysmex XE-5000: is it time for new reference ranges?
Linda M Sandhaus1, Pedro Ciarlini, Diane Kidric
1Dept of Pathology, University Hospitals Case Medical Center, 11100 Euclid Ave, Cleveland, OH 44106, USA.
American Journal of Clinical Pathology
|October 21, 2010
Summary
Automated white blood cell (WBC) counts on cerebrospinal fluid (CSF) show weaker correlation at low cell levels compared to manual methods. Further research is needed to establish reliable automated reference ranges for CSF WBC analysis.
Area of Science:
- Clinical Laboratory Science
- Hematology
- Neuroscience
Background:
- Accurate white blood cell (WBC) counts in cerebrospinal fluid (CSF) are crucial for diagnosing neurological conditions.
- Manual cell counting using a hemocytometer is the traditional method, but it is labor-intensive.
- Automated methods offer potential for faster and more efficient CSF analysis.
Purpose of the Study:
- To compare the accuracy of manual WBC counts with automated counts from the Sysmex XE-5000.
- To evaluate the performance of automated CSF WBC counting across different cell concentration ranges.
- To identify potential discrepancies in automated WBC counts, particularly at low cell levels.
Main Methods:
- Analysis of 200 clear cerebrospinal fluid (CSF) samples from adult and pediatric patients.
- Manual WBC counts performed using a hemocytometer.
- Automated WBC counts performed using the Sysmex XE-5000 automated hematology analyzer.
- Method comparison using descriptive statistics and Spearman rank correlation.
Main Results:
- Manual WBC counts ranged from 0 to 702 cells/μL; Sysmex counts ranged from 0 to 629 cells/μL.
- Overall Spearman rank correlation coefficient was 0.77 (P < .001), indicating good agreement.
- Correlation was notably weaker at lower cell concentrations.
- For manual counts of 0-5 cells/μL, Sysmex 95th percentile was 0-23 cells/μL; for 0-10 cells/μL, Sysmex 95th percentile was 0-27 cells/μL.
Conclusions:
- Automated WBC counting on clear CSF samples demonstrates good overall correlation with manual methods.
- Discrepancies are more pronounced at low cell counts, suggesting limitations for precise quantification in this range.
- Larger studies are required to establish new, validated reference ranges for automated CSF WBC counts.