Introduction:
Measuring breathing effort during sleep with an oesophageal pressure sensor remains technically challenging and has not become routine practice. The aim of the present work was to investigate whether increased thoracic pressure during sleep can be detected with the Emfit movement sensor. Experimental data suggest that increased respiratory efforts with the intrathoracic pressure variation induce high-frequency spikes in the Emfit signal, but this has not been systematically examined.
Methods:
Polysomnography, oesophageal pressure and Emfit signal were recorded in 32 patients with suspected sleep-disordered breathing. Increased respiratory effort was defined as oesophageal pressure below -8 cmH(2)O during inspiration. The epochs of normal breathing, periodic breathing patterns and sustained spiking labelled as increased respiratory resistance (IRR) were defined on the Emfit signal according to established rules.
Results:
Compared to normal breathing, the proportion of increased respiratory effort was higher during all periodic breathing with spiking. The highest proportion (18-23%) occurred during IRR, which is characterised by sustained spiking.
Conclusion:
The Emfit movement sensor is a non-invasive alternative to the oesophageal pressure sensor in the assessment of the respiratory effort during sleep. In particular, the Emfit sensor enhances detection of non-apnoeic sleep-disordered breathing, the significance of which should not be ignored.
More Related Videos
Related Concept Videos
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Factors Affecting Respiration
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like asthma or COPD,...
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration can...
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes include...