Determination of daptomycin in human plasma by liquid chromatography-tandem mass spectrometry. Clinical application
Marie-Clémence Verdier1, Danièle Bentué-Ferrer, Olivier Tribut
1Biological Pharmacology and Toxicology, University Hospital, Rennes 1 University, Rennes, France.
Background:
Daptomycin is a recently developed cyclic lipopeptide antibiotic active against most Gram-positive pathogens including vancomycin-resistant enterococci and methicillin-resistant Staphylococcus aureus. To optimize treatment efficacy and safety, especially in patients undergoing multiple drug regimens and/or co-morbidities, a specific liquid chromatography-tandem mass spectrometry (LC-MS/MS) method was developed for the quantification of daptomycin in plasma.
Methods:
A C18 column was used for separation, with a mobile phase initially consisting of 0.1% formic acid, water, and acetonitrile (ACN) in a linear gradient from 20% to 70%. After protein precipitation with ACN, the clear upper layer was diluted in water:ACN (50:50, v/v) before injection. Detection was performed using an electrospray ionization technique. MS/MS transitions, monitored in the positive ion mode were m/z 811.1 → m/z 313.1 for daptomycin, and m/z 609.4 → m/z 194.9 for reserpine, used as internal standard.
Results:
Elution of daptomycin and reserpine occurred at 4.5 and 3.9 min, respectively. The method was validated over a range of concentrations from 1 mg/L to 120 mg/L. The assay met recommended acceptance criteria: coefficients of variation were <6.3% and <7.4%, and accuracies were between -5.9% and +11.2% and between -3.5% and +3.7%, for intra- and inter-day validations, respectively.
Conclusions:
This method appears well-adapted to routine hospital practice for therapeutic drug monitoring of daptomycin considering its time of analysis, range of concentrations measured, precision and accuracy.
Related Concept Videos
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Therapeutic Drug Monitoring: Drug Analysis Methods
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
MALDI-TOF Mass Spectrometry
Rapid Identification of Pathogens
Protein-Drug Binding: Determination Methods
Indirect methods involve isolating the bound drug from its free form in biological samples such as blood, serum, or plasma. These techniques aim to measure the percentage of drugs bound to proteins. Equilibrium dialysis is a commonly used method where the free drug concentration at equilibrium is measured by separating the bound...