Related Experiment Video
Updated: Jun 7, 2026

Fully Endoscopic Mitral Valve Repair with Percutaneous Cannulation of Groin Vessels
Published on: May 26, 2023
Pre-stenting with a bare metal stent before percutaneous pulmonary valve implantation: acute and 1-year outcomes
Johannes Nordmeyer1, Philipp Lurz, Sachin Khambadkone
1Cardiovascular Unit, UCL Institute of Child Health, London, UK.
Objectives:
To determine the feasibility and safety of pre-stenting with a bare metal stent (BMS) before percutaneous pulmonary valve implantation (PPVI), and to analyse whether this approach improves haemodynamic outcomes and impacts on the incidence of PPVI stent fractures.
Design:
Retrospective analysis of prospectively collected data.
Setting:
Tertiary paediatric and adult congenital heart cardiac centre.
Patients And Interventions:
108 consecutive patients with congenital heart disease underwent PPVI between September 2005 and June 2008 (54 with PPVI alone, 54 with BMS pre-stenting before PPVI).
Results:
There were no significant differences in procedural complication rates. Acutely, there was no difference in haemodynamic outcomes. Serial echocardiography revealed that in the subgroups of 'moderate' (26-40 mm Hg) and 'severe' (>40 mm Hg) right ventricular outflow tract (RVOT) obstruction, patients with pre-stenting showed a tendency towards lower peak RVOT velocities compared to patients after PPVI alone (p=0.01 and p=0.045, respectively). The incidence of PPVI stent fractures was not statistically different between treatment groups at 1 year (PPVI 31% vs BMS+PPVI 18%; p=0.16). However, pre-stenting with BMS was associated with a lower risk of developing PPVI stent fractures (HR 0.35, 95% CI 0.14 to 0.87, p=0.024). The probability of freedom from serious adverse follow-up events (death, device explantation, repeat PPVI) was not statistically different at 1 year (PPVI 92% vs BMS+PPVI 94%; p=0.44).
Conclusions:
Pre-stenting with BMS before PPVI is a feasible and safe modification of the established implantation protocol. Pre-stenting is associated with a reduced risk of developing PPVI stent fractures.