Parameters affecting radiographic contrast.

N L Frederiksen1, P W Goaz

  • 1Baylor College of Dentistry, Dallas, Texas.

Dento Maxillo Facial Radiology
|November 1, 1990
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Computerized tomography-based imaging and surgical guidance in oral implantology.

The Journal of oral implantology·2006
Same author

Lingual vascular canals of the interforaminal region of the mandible: evaluation with conventional tomography.

Dento maxillo facial radiology·2004
Same author

Focal trough and patient positioning.

Dento maxillo facial radiology·2004
Same author

Intraosseous oral leiomyoma: systematic review and report of one case.

Dento maxillo facial radiology·2004
Same author

Multiple idiopathic cervical root resorption: systematic review and report of four cases.

Dento maxillo facial radiology·2003
Same author

Pneumatization of the pterygoid process of the sphenoid bone.

Dento maxillo facial radiology·2001
Same journal

Cranial Base Characteristics in Adults with Obstructive Sleep Apnoea and Matched Controls: A Retrospective CBCT Study.

Dento maxillo facial radiology·2026
Same journal

Dual-Circle: a new way of assessing Temporomandibular joint (TMJ) disc position at coronal MRI.

Dento maxillo facial radiology·2026
Same journal

Quantification of Supracrestal Tissue Attachment Using High-Frequency Ultrasound.

Dento maxillo facial radiology·2026
Same journal

Radiological-pathological correlation of tumour size and depth of invasion of oral squamous cell carcinoma using Dual-Energy CT in comparison to MRI and the impact on T-stage.

Dento maxillo facial radiology·2026
Same journal

SinusNet+: Deep Condition-Label-Free Segmentation of Maxillary Sinus Conditions in CBCT images.

Dento maxillo facial radiology·2026
Same journal

Handheld dental X-ray devices: a scoping review of radiation safety and image quality.

Dento maxillo facial radiology·2026
See all related articles

Image contrast in X-ray imaging is often linked to kilovoltage and half-value layer. However, this study found no correlation with kilovoltage, but a strong link between half-value layer and image contrast.

Area of Science:

  • Medical Imaging
  • Radiology
  • X-ray Physics

Background:

  • Kilovoltage and half-value layer are commonly considered key factors for defining X-ray image contrast.
  • This has led to the assumption that consistent image contrast can be achieved by operating X-ray units at the same kilovoltage and half-value layer.

Purpose of the Study:

  • To evaluate the validity of the assumption that consistent kilovoltage and half-value layer settings yield comparable image contrast.
  • To investigate the relationship between X-ray beam characteristics and resultant image contrast.

Main Methods:

  • Seven X-ray units were tested, comparing their generated X-ray beams.
  • Image contrast was assessed after X-ray beams passed through aluminum absorbers.
  • Comparisons were made against a Gendex Model 1000 unit operated at consistent kilovoltage settings.

Related Experiment Videos

Main Results:

  • No correlation was found between the nominal operating kilovoltage of X-ray generators and the produced image contrast.
  • A strong negative correlation (-0.98) was observed between the measured half-value layer and image contrast.
  • The measured half-value layer proved to be a more reliable indicator of image contrast than the stated kilovoltage.

Conclusions:

  • The findings challenge the conventional reliance on kilovoltage for predicting X-ray image contrast.
  • Measured half-value layer is a critical parameter for achieving consistent image contrast.
  • This information should influence the selection and operational criteria for X-ray generators to optimize image quality.