Related Experiment Video
Updated: Jun 7, 2026

Sample Preparation and Transfer Protocol for In-Vacuum Long-Wavelength Crystallography on Beamline I23 at Diamond Light Source
Published on: April 23, 2021
Development of diamond-based X-ray detection for high-flux beamline diagnostics
Jen Bohon1, Erik Muller, John Smedley
1Center for Synchrotron Biosciences, Case Western Reserve University, Upton, NY 11973, USA. jbohon@bnl.gov
Abstract:
High-quality single-crystal and polycrystalline chemical-vapor-deposition diamond detectors with platinum contacts have been tested at the white-beam X28C beamline at the National Synchrotron Light Source under high-flux conditions. The voltage dependence of these devices has been measured under both DC and pulsed-bias conditions, establishing the presence or absence of photoconductive gain in each device. Linear response consistent with the theoretically determined ionization energy has been achieved over eleven orders of magnitude when combined with previous low-flux studies. Temporal measurements with single-crystal diamond detectors have resolved the nanosecond-scale pulse structures of both the NSLS and the APS. Prototype single-crystal quadrant detectors have provided the ability to simultaneously resolve the X-ray beam position and obtain a quantitative measurement of the flux.
Related Concept Videos
X-ray Diffraction of Biological Samples
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal crystal...
Determination of Crystal Structures