Related Experiment Video
Updated: Jun 7, 2026

Activated Cross-linked Agarose for the Rapid Development of Affinity Chromatography Resins - Antibody Capture as a Case Study
Published on: August 16, 2019
Reversible interference of Fe³(+) with monoclonal antibody analysis in cation exchange columns
1R&D, Dionex Corporation, Sunnyvale, CA 94085, United States. srinivasa.rao@dionex.com
Abstract:
The effect of Fe³(+) treatment on weak cation exchange column chromatography was demonstrated for monoclonal antibody (MAb) analysis. Fe³(+)-exposed columns showed lowered relative peak areas of the parent MAb peak as well as both acidic and basic variant peaks that could lead to erroneous conclusions. Accurate measurement of relative amounts of variants to the parent MAb is essential for demonstrating the safety and efficacy of therapeutic molecules such as MAbs. Complete reversal of the compromised MAb analysis performance was observed after washing the column with chelating agents, confirming that metal contamination was responsible for the compromised column performance.
More Related Videos
Related Concept Videos
Ion-Exchange Chromatography
Ion Exchange
Affinity Chromatography
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences