Optimization of high-resolution USPIO magnetic resonance imaging at 4.7 T using novel phantom with minimal structural
Henrik Antell1, Jussi Numminen, Usama Abo-Ramadan
1Biomedicum Helsinki, Neurosurgery Research Group, Helsinki, Finland. henrik.antell@hus.fi
Purpose:
To characterize the effect of ultrasmall superparamagnetic iron oxides (USPIOs) on magnetic resonance imaging (MRI) signal at 4.7 T, and to find the highest sensitivity pulse sequence for high-resolution USPIO MRI.
Materials And Methods:
A novel phantom was constructed for optimization of sequence parameters for neuroradiological MR applications, and a wide range of dilutions of the USPIO ferumoxtran-10 was imaged using T(2)/T(1)-, T(1)-, T(2)-, T* (2)-, and PD-weighted sequences. The effect of varying sequence parameters was investigated using phantom measurements and simulations.
Results:
The relaxivities r(1), r(2), and r*(2) of ferumoxtran-10 at 4.7 T (21°C) were 5.1, 82.2, and 148.4 mmol(-1) L s(-1), respectively. Gradient echo sequences produced superior susceptibility artifacts at high concentrations; susceptibility artifacts were seen down to a concentration of 137 nmol Fe/mL. A concentration of 17.5 μmol Fe/mL caused a signal void independently of sequence and parameters, and at concentrations ≤273 nmol Fe/mL no signal void was caused. Signal enhancement on T(1)-weighted imaging was seen only at concentrations 137-547 nmol Fe/mL. For the same effective echo time T(2)-weighted rapid acquisition with relaxation enhancement (RARE) yielded significantly higher contrast-to-noise ratio with RARE factor 16 than with RARE factor 8.
Conclusion:
At nanomolar concentrations of USPIO, steady-state free precession offers an alternative to T(2)- and T*(2)-weighted sequences. Optimum parameters depend highly on USPIO concentration.