Development of an x-ray imaging system for the Laser Megajoule (LMJ)
G Turk1, C Reverdin, D Gontier
1CEA, DAM, DIF, F-91297 Arpajon, France. gregory.turk@cea.fr
Abstract:
This imaging system aims at recording images of the core size and shape of an imploding deuterium-tritium (DT) microballoon on LMJ inertial confinement fusion (ICF) experiments. Image acquisition is difficult due to the harsh surrounding created by the fusion reaction, which affects system specifications. This one is made of a scintillator, an optical relay, and a CCD camera shielded from the surrounding. The system was tested on different facilities at CEA/DIF, where a spatial resolution of 120 μm was achieved and gamma dose up to 20 rad effects were measured. Setup and performed test are described.
More Related Videos
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...