Related Experiment Video
Updated: Jun 7, 2026

The Frequency Domain Thermoreflectance Technique for Thermal Property Measurements
Published on: December 5, 2025
Emissivity measurements of reflective surfaces at near-millimeter wavelengths
Abstract:
We have developed an instrument for directly measuring the emissivity of reflective surfaces at near-millimeter wavelengths. The thermal emission of a test sample is compared with that of a reference surface, allowing the emissivity of the sample to be determined without heating. The emissivity of the reference surface is determined by one's heating the reference surface and measuring the increase in emission. The instrument has an absolute accuracy of Δε = 5 × 10(-4) and can reproducibly measure a difference in emissivity as small as Δε = 10(-4) between flat reflective samples. We have used the instrument to measure the emissivity of metal films evaporated on glass and carbon fiber-reinforced plastic composite surfaces. We measure an emissivity of (2.15 ± 0.4) × 10(-3) for gold evaporated on glass and (2.65 ± 0.5) × 10(-3) for aluminum evaporated on carbon fiber-reinforced plastic composite.
Related Concept Videos
Interaction of EM Radiation with Matter: Spectroscopy
Attenuated Total Reflectance (ATR) Infrared Spectroscopy: Overview
The ATR process begins by directing a beam...
Spectrophotometry: Introduction
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
UV–Vis Spectrometers
IR Spectrometers