Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024
How reliable is visual assessment of the electrically elicited stapedius reflex threshold during cochlear implant
1Department of Otorhinolaryngology, Head and Neck Surgery, University of Rostock, Rostock, Germany. hans-wilhelm.pau@med.uni-rostock.de
Objectives:
To assess the reliability of visually assessed thresholds of the electrically elicited stapedius reflex, recorded during cochlear implant surgery, compared with intra-operative tympanometric threshold assessment. Intra-operatively recorded electrically elicited stapedius reflex thresholds vary considerably, and differ from those measured post-operatively by means of impedance changes (i.e. using tympanometry). Thus, any confounding effect of different intra-operative techniques and visual assessment inaccuracies should be excluded.
Methods:
Both techniques (i.e. visual observation and tympanometry) were performed intra-operatively in six patients, and threshold values were compared.
Results:
Recorded electrically elicited stapedius reflex thresholds were very similar for both techniques. Visually assessed thresholds were slightly higher in some cases and lower in others, compared with tympanometric thresholds.
Discussion:
There was almost no difference between reflex thresholds measured with the two different techniques under the same intra-operative conditions. Therefore, we conclude that differences between intra- and post-operative thresholds are not due to the use of different measuring techniques. The main reason for such differences is probably the influence of intra-operative narcotics on reflex thresholds.